Probe Compensation - Owon VDS6000 Series Quick Manual

Hide thumbs Also See for VDS6000 Series:
Table of Contents

Advertisement

7. Probe Compensation

Before working the probe with either of input channels (Channel 1 / Channel 2), better to
adjust its compensation, to assure ideal measurement effect. Since the probe in
non-compensated, or improperly-compensated (overcompensated / under-compensated)
status always possibly cause error or mistake of measurement result.
Following operation steps to adjust probe compensation -
i. From PC software operation interface, mouse-click
function menu, choose "Channel", then "CH1", set the "Probe Rate" as "x10". Switch
probe attenuation to X10 option, then get the probe through Channel 1 input via device
communication interface. Provided introducing a probe hook tip, make sure it is in close
touch with the probe. Connect the probe tip with the signal output coupler of probe
compensator, and connect the reference wire clamp with the ground wire coupler of the
probe compensator, and then click
Figure 7- 1. Probe Attenuation Switch Position (could set in 1X or 10X option)
ii. The probe compensation goes in 3 different status (as Figure 7- 2 shows),
Overcompensated
provided probe in overcompensated or under-compensated status, use probe adjust (from
device accessories) to adjust the screw (leftwards or rightwards) of probe body (Figure 7- 3).
For different brand-new probe, please repeat these steps to get ideal compensation.
to do auto set.
Under-compensated
Figure 7- 2. Different Status of Probe Compensation
Figure 7- 3. Probe Screw Adjusting
- 16 -
to get access to device
Well-compensated

Advertisement

Table of Contents
loading

Table of Contents