Hach Lange astroTOC Basic User Manual page 22

Uv/turbo analyzer

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Startup
Reagent preparation
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current material safety data
sheets (MSDS) for safety protocols.
Refer to the CD for reagent preparation information.
Plumb the reagent solutions
Install the reagent solutions as shown in
Figure 14 Reagent connections
1 Tubing assembly (4 L)
2 Tubing assembly (19 L)
Turn on the analyzer
Turn on the main AC power to the analyzer.
After a purge, the pumps and UV lamps turn on and the IR detector increases to operating
temperature. The analyzer goes online when the IR detector is at operating temperature (50 °C).
After a loss of power or a software reboot, the analyzer turns on and a purge is done before the
analyzer goes online. To skip the purge, push ENTER.
Start-up checks
1. When the display shows "Purging", make sure that the analyzer has the correct range for the
intended application.
2. Make sure that the sample inlet line(s) is in the zero standard container.
22 English
W A R N I N G
Figure
14.
3 Span solution
4 Zero solution
5 Persulfate solution
6 Acid solution