Hach TSS Portable Troubleshooting Manual
Hide thumbs Also See for TSS Portable:

Advertisement

TSS Portable Troubleshooting Guide

TABLE OF CONTENTS

Scope and Principle ............................................................................................................... 2
Presale Information ............................................................................................................... 2
Accessories/Replacement Parts ............................................................................................. 2
Product Contents................................................................................................................... 4
Software ............................................................................................................................... 6
Calibration .......................................................................................................................... 14
Measurement ...................................................................................................................... 15
Maintenance ....................................................................................................................... 16
FAQ ..................................................................................................................................... 17

Advertisement

Table of Contents
loading

Summary of Contents for Hach TSS Portable

  • Page 1: Table Of Contents

    TSS Portable Troubleshooting Guide TABLE OF CONTENTS Scope and Principle ....................... 2 Presale Information ....................... 2 Accessories/Replacement Parts ..................... 2 Product Contents........................4 Software ..........................6 Calibration .......................... 14 Measurement ........................15 Maintenance ........................16 FAQ ............................. 17...
  • Page 2: Scope And Principle

    Drinking water: Spring Water, Drinking Water Reservoir, Filtering Station, Ground Water Pumping Presale Information 1. Only TSS portable instrument on the market that can measure all three parameters: suspended solids, sludge blanket level, and turbidity. 2. Ability to store 1 turbidity calibration curve and 4 calibration curves for suspended solids a.
  • Page 3 2. Probe Only - LZV321.99.00001 3. Meter Only - LXV320.99.00001 4. Rechargeable NiMH Batteries - LZY604 5. Battery Holder - LZY606 6. Charger - LZY607 7. Protective Beaker - LZI-66.11487...
  • Page 4: Product Contents

    Product Contents 1. Complete kit: LXV322.99.00002 2. TSS portable handheld instrument 3. Probe with 10m cable 4. Adapter plugs: a. Charger with EU adapter plug installed, USA adapter plug, UK adapter plug, AUS/China adapter plug 5. Carrying Case 6. Quick Start guide (DOC013.53.90141)
  • Page 5 RECOMMENDED CALIBRATION EQUIPMENT 1. Black container (dark bucket) To avoid reflections of the measurement beam on the base and walls during calibration. 2. Container should hold at least 1L of sample 3. The probe must swirl inside the bucket and have a clearance of 2 inches from the side walls and 3 inches from the bottom of the container.
  • Page 6: Software

    4. To go back one screen: Press the CLEAR OFF key 5. Entering multi-digit numbers: Use left or right key navigation keys to move curser to the correct place (1,10,100,1000nds). Then use the up or down navigation keys to select the desired number.
  • Page 7 b. The measurement is continuous SOFTWARE MENU 1. Main Menu structure overview 2. Display Menu...
  • Page 8 3. Menu Structure “MEAS” with function a. Use up and down Navigation key to move through the “MEAS” menu...
  • Page 9 4. Menu structure “DATA” with function a. Display Data b. Delete Data: *YES will delete all data points for that curve...
  • Page 10 5. Menu structure “MENU” with function a. “Calibration” menu with function...
  • Page 11 b. “Time/Date” menu with function c. “Display” menu with function...
  • Page 12 d. “Units” menu with function e. “Extras” menu with function...
  • Page 13 “Language” and “System” menu with function SOFTWARE ERROR MESSAGES 1. Error Messages: As taken from the manual 3 edition section 7 a. MEAS OVERRANGE: Concentration too high Action: Medium cannot be measured b. ZERO: Calibration pint too close to zero Action: Delete calibration and start again...
  • Page 14: Calibration

    c. LED FAILURE: Faulty LED Action: Send the probe for repair d. EE RSRVD ERR/EE SETUP ERR: Meter fault Action: Send the meter for repair e. ACCU FLAT: old software version for a low battery 2. Informative Messages: Calibration (Turbidity and Suspended Solids) 1.
  • Page 15: Measurement

    6. Select MEMORY, ENTER, and the desired POINT. Once the point is selected and ENTER is pressed the probe will start to measure the sample. 7. Stir the probe in the sample at a steady rate. “Wait” is visible with MEMORY on the display. 8.
  • Page 16: Maintenance

    c. Select DATA, ENTER, DISPLAY DATA, ENTER d. Use the navigation key UP/DOWN to select the desired point e. Selecting DATA. ENTER, DELETE DATA, YES will delete all data for that curve. SLUDGE BLANKET LEVEL 1. Select appropriate suspended solids curve 2.
  • Page 17: Faq

    This expected life will shorten as the batteries age. Q: Is the there a power cord available for the TSS portable system? A: No, the TSS portable system will only come with a battery charger. It is recommended to turn the meter off while charging the batteries.
  • Page 18 A: No, the meter does not have a way to connect to a storage device or a computer. Q: What do I do if Hach changes the software version of the meter? A: The meter will have to be sent to the appropriate service center for the software update.
  • Page 19 Setting up a new calibration curve is recommended. If the problem continues, the sludge concentration is too high to be measured with the TSS portable probe; the value is beyond the probes measurement range.
  • Page 20 Q: Is there general guidance for creating the most accurate calibration curve for suspended solids? A: Hach recommends selecting a calibration point with a value in the upper third of the range expected. Start with a single point calibration in that range. In case a single point calibration is not sufficient, the error “message Calibration insufficient –“or “Calibration insufficient +”...
  • Page 21 % of homogeneity is high. The ratio between water and sludge and the structure of the sludge is responsible for a high homogeneity value. Q: Will the TSS portable system work correctly on low or not fully charged batteries? A: Customers have reported issues with taking measurements, frozen menus, meters powering off unexpectedly, and calibration issues when the batteries were low or not fully charged.