Section 1 Product overview The Intellical LDO101 series probes are digital, luminescent dissolved oxygen probes that measure the dissolved oxygen concentration in wastewater, drinking water and general water samples. The probes have temperature and absolute air pressure sensors for accurate dissolved oxygen measurements.
Specifications Details Dissolved oxygen accuracy ±0.1 mg/L for concentrations less than 8 mg/L ±0.2 mg/L for concentrations more than 8 mg/L % saturation resolution 0.1% Stabilization time T90% at 10 seconds (when the solution is stirred) Temperature resolution 0.1 °C (0.18 °F) Temperature accuracy ±0.3 °C (±0.54 °F) Pressure resolution...
N O T I C E Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis. 3.3 Precautionary labels Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed.
Figure 2 Compare lot number Section 5 Calibration The procedure that follows is applicable to meters that can connect to Intellical LDO probes. Refer to the applicable meter documentation for meter operation and probe-specific settings. 5.1 Calibration notes Read the notes that follow before calibration. •...
5.2 LDO water-saturated air (100%) calibration procedure 1. Fill a BOD bottle 2. Put a stopper on 3. Rinse the probe 4. Put the probe in ¼ full with water. the bottle and shake with deionized water. the bottle. Wait for 30 seconds.
• Make sure to install the shroud before field use to prevent damage to the sensing elements. Refer Remove or install the shroud on page 9. The probe warranty does not include such damage. • To deploy a rugged probe at a distance, toss the probe body with a slow underhand throw. Do not throw the probe by the cable to prevent damage to cable, the probe or the user.
Figure 3 Probe exploded view 1 Shroud 7 Probe lens 2 Locking ridges (8x) 8 Cap seal 3 Locking rib 9 Temperature sensor 4 Locking ring 10 Locking groove 5 LDO sensor cap 11 Locking ribs (4x) 6 O-ring 7.2 Replace the sensor cap and iButton Replace the LDO sensor cap after 365 days, or more frequently if the cap becomes damaged or dirty.
Figure 5 Install the shroud 7.4 Storage The storage conditions for the LDO probe can be dry, wet, or humid. Wet storage conditions help to keep the LDO sensor cap fully hydrated. Note: After 72 hours of storage in tap water, the LDO sensor cap is fully hydrated. For daily measurements in the laboratory, keep the probe in a humid environment such as a BOD bottle filled approximately ¼...
Section 8 Troubleshooting Problem Possible cause Solution Decreased probe performance The sensor cap is loose or Tighten or replace the sensor cap. causes slow stabilization and damaged. Always replace the sensor cap and prevents accurate calibrations the iButton at the same time. Refer or measurements.
Page 15
Section 9 Consumables (continued) Description Quantity Item no. Wash bottle, polyethylene, 500 mL 62011 Shroud kit for standard LDO probes 5832500 Shroud kit for rugged probes 5825900 English 13...