Installation
2
40
Cubis
®
MCA Operating Instructions
5.6
Procedure
Turn the device on its side and place it on the soft support base .
t
Unscrew both screws (2) from beneath the weighing module with the
t
Torx Allen key .
1
Lift out the cover plate (1) .
t
Install a draft protection shield .
t
Set up the device on the weighing table with recess . The hook for below-balance
t
weighing (3) may not touch the weighing table .
Install the draft protection shield .
t
3
5.7
Installing a Microbalance with Glass Draft Shield
5.7.1
Positioning the Weighing Pan and Associated Components
Procedure
For a device with a connector: Place the connector (2) on the base of the
t
weighing chamber .
Place the shield plate (1) on the base of the weighing chamber .
t
1
2