Power Supply Wiring Connections

Connect power to the screw terminal block marked J2 through the conduit hole on the left side of the enclosure See
for AC power supplies and
example, The National Electric Code Handbook in the U S )
CAUTION
ANY OTHER WIRING METHOD MAY BE UNSAFE OR CAUSE IMPROPER OPERATION OF THE INSTRUMENT.
TO AVOID SERIOUS INJURY OR DAMAGE, DISCONNECT ELECTRICAL POWER BEFORE SERVICING THIS METER.
N
OTEE:
This instrument requires clean electrical line power Do not operate this unit on circuits with noisy components (such
as fluorescent lights, relays, compressors or variable frequency drives) Do not run line power with other signal wires
within the same wiring tray or conduit
August 2019
Figure 16
for DC power supplies Use wiring practices that conform to local and national codes (for
WIRING DIAGRAM
To avoid serious injury or damage,
CAUTION!
disconnect electrical power before servicing this meter
JP3
Connections
2 1
115 VAC
4 3
2 1
230 VAC
4 3
2 1
9-28 VDC
4 3
JP1/JP2
Connections
2 1
115/230
VAC
2 1
2 1
9-28 VDC
2 1
AC
L1
L2
DC
+V
GND
SYNC SELECT
Figure 15: DFX flow meter wiring diagram
DPP-UM-00256-EN-03
JP3
2 1
4 3
1
JP2
JP1
2 1
J4
J3
J2
Fuse (5x20mm)
AC: 0.25 A/250V Delay
DC: 1.0 A/250V Delay
RED BLK
BLU CLR
EARTH
Receive Transmit
EARTH
EXT SYNC
INT
EXT
Transmitter Installation
MODULE #2
MODULE #1
BLK
RED
CLR
BLU
GND
Figure 15
Page 21