2 6
Maintenance of the instrument
To obtain consistently exact measurement results please follow these
instructions for cleaning and maintenance.
Handling
1 . The instrument is very robust. Nevertheless you should handle it carefully
as with any other electronic device.
2 . Do not expose the instrument to excessive humidity and heat or cold
(see Technical data).
3 . The outside case and the display can be cleaned with a moist cloth.
Do not use solvents or abrasive cleaning agents.
Cleaning the strip adapter
When?
The strip adapter must be cleaned immediately if an error message
(OPT or ERR) is displayed.
The strip adapter should be thoroughly cleaned (dismantling the strip
adapter into its individual components) on a regular basis. Please refer to
the Reflectoquant® insert for the recommended period of use.
How?
1 . Switch off the instrument.
2 . Carefully pull the strip adapter out of the case.
3 . Dismantle the adapter into its three components.
4 . Clean the components with water and a mild detergent.
If necessary use ethanol. NEVER clean the white section
with abrasive cleaner as this will destroy its reflective properties.
5 . Dry the components carefully and reassemble the adapter.
6 . Reinsert the adapter into the instrument.