Features - BRASSELER USA EndoSync TR-CM BSL Operation Instructions Manual

Cordless endodontic treatment motorized handpiece
Table of Contents

Advertisement

Features

⹅ Indications for Use
The EndoSync is a compact, cordless endodontic treatment motorized handpiece for preparation and enlargement of root canals.
It may be connected to the EndoSync A.I., an apex locator (sold separately).
It can be used to enlarge and prepare root canals, remove of gutta-percha point and softened dentin, and professional mechani-
cal tooth cleaning.
Instructions for how to use the EndoSync when it is connected to the EndoSync A.I. are printed on a blue background like this one.
⹅ Liquid Crystal Display (LCD) :
The LCD is easy to read and shows all settings as well as how the motor is running.
⹅ Controls :
< OTR (Optimum Torque Reverse) mode >
If the file torque is less than the set value, the file will keep rotating in the forward direction.
When the file torque is more than the set value, the file will automatically start rotating 90° in reverse and 180° forward repeat-
edly. Furthermore, the OTR mode can set various motor controls as described below.
Speed :
Torque Setting :
Slow Down :
Auto Start & Stop* :
Apical Reverse or Stop* :
Apical Torque Reduction* :
< Normal Mode >
If the file torque is less than the set value, the file will keep rotating in the forward direction. When the file torque is more than
the set value, the file will automatically start rotating in reverse direction.
Furthermore, the Normal mode can set various motor controls as described below.
Speed :
Torque Reverse :
Slow Down :
Forward & Reverse :
Auto Start & Stop* :
Apical Reverse or Stop* :
Apical Torque Reduction* :
⹅ Memory:
Six combinations of speed, torque etc. can be memorized.
Operation 2016-03-07
100, 300, 500 rpm.
0.2, 0.4, 0.6, 0.8, 1.0 Ncm.
The file slows down as torque increases.
The file slows down as it approaches the apex if the EndoSync is connected to the EndoSync
A.I.
The file starts when it is inserted in the canal and stops when it is taken out.
The motor reverses or stops when the tip of the file reaches a preset position inside the ca-
nal.
The automatic torque reverse value is reduced as the file tip approaches the apex.
* These controls can be used if the EndoSync is connected to the EndoSync A.I.
Eleven rotation speeds can be set from 50 to 1,000 rpm.
The motor automatically reverses its rotation if the torque load exceeds the set value to re-
duces the risk of jamming.
The file slows down as torque increases.
The file slows down as it approaches the apex if the EndoSync is connected to the EndoSync
A.I.
The file may rotate in both forward and reverse directions.
The file starts when it is inserted in the canal and stops when it is taken out.
The motor reverses or stops when the tip of the file reaches a preset position inside the ca-
nal.
The automatic torque reverse value is reduced as the file tip approaches the apex.
* These controls can be used if the EndoSync is connected to the EndoSync A.I.