File Installation; Calibration - BRASSELER USA EndoSync TR-CM BSL Operation Instructions Manual

Cordless endodontic treatment motorized handpiece
Table of Contents

Advertisement

File Installation

File Release Button
2
1

Calibration

* Before using right after purchase, whenever the motor handpiece or contra angle has been replaced, or if the motor alternates
between forward and reverse rotation outside the canal, calibrate the instrument in the following way:
• Never use deformed or damaged files.
• Give the file a light tug to confirm it is securely held in place. If the file is not securely placed, it could come out and injure the patient.
• Do not use reciprocal files (ones made to rotate back and forth). These could perforate the apical foramen when they reverse rotation.
• Use caution when inserting and removing files to avoid injury to fingers.
• Inserting and removing files without holding the file release button down will damage the chuck.
• Make sure the EndoSync is turned OFF before inserting or removing files.
• Do not connect the file electrode if the motor handpiece is not connected to the EndoSync A.I.
Operation 2016-03-07
3
1. Make sure the battery is fully charged (three bars are displayed).
2. Put a commonly used file into the contra angle.
3. Turn the unit ON and select memory M6.
4. Set the rotation mode for Rev.
5. Turn the unit OFF .
6. Hold down the plus and minus switches and turn the unit back ON.
7 . W h e n " C O - A d j s t " a p p e a r s i n t h e d i s p l a y, p r e s s t h e S e l e c t ( S ) s w i t ch .
The motor will start running. Make sure there is no load on the file.
8. When "Finished" appears in the display, the motor will stop and the calibration will be com-
pleted.
Press the main switch to go to the standby display.
* Once calibration has been completed, you may change the M6 setting and you may turn the
unit OFF with a setting other than M6.
* If the unit uses a file electrode, calibrate the instrument by putting the electrode on the file
and connecting the unit to a EndoSync A.I. unit which is turned ON.
Hold down the file release button
Insert the file and turn it back and forth until it lines up
with the latch mechanism.
Push the file all the way into the latch. Release the file
release button.
It's easiest to hold the button down if you put your index fin-
ger where the head joints the body.
Use either Nickel-titanium or stainless steel files.
10

Advertisement

Table of Contents
loading

Table of Contents