Connect Transmission Cable; Check Operation - BRASSELER USA EndoSync TR-CM BSL Operation Instructions Manual

Cordless endodontic treatment motorized handpiece
Table of Contents

Advertisement

Connect Transmission Cable

* Refer to the user manual for the EndoSync A.I.
1
Plug: White
Take off the cap and plug the cable's white plug all
the way into the motor handpiece.
Do not mix up the cable plugs.

Check Operation

■ Make sure file electrode is making good contact with the file.
■ Touch the file with the contrary electrode and make sure the meter goes all the way to its end and there are no seg-
ments that do not light up.
Watch out because the motor might start up when you do this.
• Use only the special cable provided. Other cable could be electrically risky and result in damage or injury. Make sure the file goes all the way in. Give the file
a light tug to make sure it is properly installed.
• Check the meter activity before each patient and do not use the instrument if all the segments of the display do not light up. This suggests that the meter can-
not make an accurate reading.
• Make sure the plugs go straight in.
• After insertion give plugs a light tug to make sure they are securely connected. Otherwise, data may not be transmitted accurately.
• Do not bump the plugs or drop anything on them when they are plugged in.
19
2
Plug: Gray
Plug the cable's gray plug into the EndoSync A.I.
Operation 2016-03-07

Advertisement

Table of Contents
loading

Table of Contents