Features - BRASSELER USA EndoDrive PLUS Operation Manual

Motor handpiece system for root canal enlargement
Table of Contents

Advertisement

• EndoDrive PLUS should not be used adjacent to or stacked with other equipment and that if adjacent or
stacked use is necessary, EndoDrive PLUS should be observed to verify normal operation in the
configuration in which it will be used.
• During rotation, the motor handpiece and the motor handpiece cord may affect computer and LAN cable. Noise
could be heard during operation near a radio receiver.
• The motor handpiece consumes electricity very slightly even when the power supply is off. In addition,
fully-charged rechargeable batteries, in general, discharge gradually over time even though it is not used. It is
recommended to recharge the batteries just before use.
• When the motor handpiece stops automatically by detecting a low battery voltage, turning on the power again
after a while may not detect the low voltage immediately. This is not a failure, but due to battery characteristics.
The voltage drop does not coincide with the remaining battery capacity.
• Recharge rechargeable batteries after they deplete as much as possible. Repeating short-time use and
subsequent recharging may shorten their operating time due to the "memory effect." Batteries may recover after
repeating complete discharge and full charge for a few times. (This product is equipped with a "Refresh Mode"
for this purpose.)
• The used Nickel Metal Hydride Batteries are recyclable, but their disposal may sometimes not be permitted by
the country. Return them to Brasseler USA.
• When disposing of the control unit and Foot Control (Option), follow the instructions of your local government for
disposal, as they contain materials which may become industrial waste.
• When discarding the contra angle head and motor handpiece, dispose of them as medical waste.
• Users are responsible for operational control, maintenance and inspection.
• Store the system in the place where the temperature is at -10-60°C (14-140°F), humidity at 10-85% RH,
atmospheric pressure at 500-1060 hPa, and the system is not subject to air with dust, sulfur, or salinity.
• When trouble is found, contact Brasseler USA.
• This product does not consider patient's age (except infants), gender, weight or nationality.
• No special training is required for this device.

1. FEATURES

• Compact design.
• Able to operate for approx. 2 continuous hours at the rated load. (Dependent upon use conditions.)
• A large LCD panel enhances ease of use.
• Programmable up to 9 memory settings.
• The auto reverse function is activated depending on the load. A wide variety of functions such as "auto reverse",
"auto stop" and "auto reverse off" are available. Combined with such functions nine different operation programs
are memorable.
• The rotation of the motor handpiece can be activated by the ON/OFF button of the motor handpiece or the Foot
Control (optional).
• The motor handpiece can be turned on and off by pressing the ON/OFF button alternatively. In addition, ON by
keep depressed and OFF by releasing it is possible.
• The motor handpiece softly starts. Since it makes a brief stop before it starts in the reverse rotation, there are no
vibration and shocks during this change over.
• The built in feedback circuit maintains rotation at a constant speed even when the load on the motor handpiece
changes.
• Has energy saving feature. The power of the motor handpiece automatically turns off when not being used for 10
minutes. (Auto power-off function)
• The contra angle heads provided for this product are all autoclavable up to 135°C (excluding the motor
handpiece).