(2) Operation

Basic Operation

(Select memory and change settings)
Main Switch
(Turn power ON and starts/stops rotation)
1. Turn EndoSync ON: Press Main switch.
The standby display will appear.
When the standby display is being shown, you can turn the EndoSync OFF by holding down the Select Switch and press-
ing the Main Switch
* The EndoSync turns itself OFF automatically if it is not used for 3 minutes (initial setting).
2. Select Memory number: Press Plus or Minus switch.
* There are six memories for various combinations of speed, torque reverse and rotation mode settings.
* The backlight will temporarily change color if changing the memory number changes anything other than the speed,
torque reverse, and rotation mode settings.
3. Start motor: Press Main switch again.
The Torque Display will appear.
* If you hold the Main Switch down when you start the motor, it will run only while the switch is held down and stop when
the switch is released.
* You can temporarily change the torque reverse setting while the motor is running by pressing the Plus or Minus switch.
(Normal mode only)
When the Apical Torque Reduction is turned ON, the torque setting cannot be temporarily changed.
* The color of the backlight changes depending on the load applied to the file.
* The backlight starts blinking when the load approaches the torque reverse setting. While the OTR is triggered, the back-
light does not blink.
4. Stop motor: Press Main switch again.
The standby display will reappear.
• Do not fail to check the new settings whenever you change the Memory Number.
• The temperature of the motor handpiece rises up to 46.0°C (114.8°F) when the ambient temperature is 40 °C (104 °F).
• When the OTR seems to be triggered too frequently, or it is triggered immediately after starting the normal rotation, increase the torque setting by one level.
Operation 2016-03-07
Plus and Minus Switches
Select Switch
Liquid Crystal Display
Standby Display (OTR mode)
Standby Display (Normal mode)
Memory Number
Rotation Mode
Battery Power
Speed Setting
OTR Torque Setting
Torque Reverse Setting
When connected to the EndoSync A.I., refer to page 20
for meter readings and operation.
Torque Display
Operation Mode
Current Torque
Torque Setting
12