Rotor Axle and Built-in Electrode Cleaning

* If the bars flicker during use, or if all the bars in the meter do not light up when the file touches the contrary electrode, clean
the rotor axle and the built-in electrode in the following way.
1
Take out the screw and then take out the built-in electrode.
3
Clean the built-in electrode with the brush.
5
1
Hold the push button and slide the guide bur straight in as
shown in the illustration. Then rotate it left and right.
• Do not bend or deform the electrode.
• Always use the guide bur and make sure it will not come out. If the guide bur is not properly fix in place, the internal contact could be bent, and then the instru-
ment might not be able to make accurate measurements or else it might malfunction.
• Do not run the motor with the guide bur inserted; this could damage the instrument.
Operation 2016-03-07
3
2
2
Put a little Ethanol for Disinfection (Ethanol 70 to 80 vol%)
on a brush and clean the rotor axle with it.
4
Blow air on the electrode to remove any remaining mois-
ture.
6
Slide the built-in electrode onto the guide bur and line up
the screw holes.
Rotor Axle