Memory Settings
The initial settings are shown below. These settings can be changed.
Setting
Speed (rpm)
Torque Reverse (Ncm)
Rotation Mode
Torque Slow Down
Apical Action
Function**
Apical Reverse or
Stop**
Auto Start or
Stop**
Apical Slow
Down**
Apical Torque
Reduction**
Memory Settings: Primary Functions
Primary Functions
: Rotation Speed, Torque Reverse, Rotation Mode
Standby Display
Rotation Mode
Speed Settings:
< OTR mode >
100, 300 and 500 rpm
< Normal mode >
50, 100, 150, 200, 250, 300, 400, 500, 600, 800 and 1000 rpm
• If the torque limit is too high, the file could jam inside the canal and break.
• The torque settings must be changed depending on the root canal condition and the file.
• If the torque reverse seems to be activated too frequently increase its value.
• When the OTR mode (motor runs back and forth continuously) seems to be triggered too frequently, or it is triggered immediately after starting the normal ro-
tation, increase the torque setting by one line. < OTR mode >
13
M1
M2
500
300
OTR
—
ON
Reverse
ON
—
—
**
These functions are available only when connected to the EndoSync A.I.
Speed
Torque Reverse
Memory
M3
M4
500
500
0.2
OFF
ON
—
Reverse
—
ON
OFF
OFF
1. Select a memory number for the standby display by press-
ing the Plus or Minus switch.
2. Press the Select Switch to choose one of the primary
functions.
3. Press the Plus or Minus switch to change the setting.
* The display will go back to the standby display if 5 sec-
onds (initial setting) elapses without a switch being
pressed.
Torque reverse values may differ somewhat depending on the mo-
tor and contra angle.
Torque Settings:
< OTR mode >
0.2, 0.4, 0.6, 0.8 and 1.0 Ncm
< Normal mode >
0.2, 0.4, 0.6, 0.8, 1.0, 1.5, 2.0, 2.5 and 3.0 Ncm
This function can also be turned OFF: TRL (torque reverse-
less).
Rotation Mode:
Normal (Fwd: Forward, Rev: Reverse), OTR
M5
M6
300
500
Normal (Forward)
OFF
OFF
—
—
—
—
Operation 2016-03-07