Cleaning; Sterilization - BRASSELER USA EndoDrive PLUS Operation Manual

Motor handpiece system for root canal enlargement
Table of Contents

Advertisement

(5) Alarm Sound Volume Control
The alarm sound volume can be controlled at the three steps of "BIG VOLUME, SMALL VOLUME and LIMITED OFF. "
1) Keep pressing both buttons of the PROGRAM key simultaneously.
2) The alarm mark on the LCD panel changes, and the sound volume changes.
3) Release the PROGRAM key when the sound volume reaches your required level and the required sound level is set up.
NOTICE
• The sound level is retained, even if the power is turned off.
• If P.12 6. (2) "Initialization of Program (Factory-set configuration)" is executed, the sound is set to "BIG VOLUME."

7. CLEANING

(1) Lubricating Contra Angle Head
• Lubricate the contra angle head only.
• Supply spray after each use and/or before each calibration
and autoclaving.
1) Screw the spray nozzle onto the spray nozzle.
2) Insert the spray nozzle into the rear part of the contra angle
head and lubricate the head for 2-3 seconds. (Fig. 12)
(2) Cleaning Motor Handpiece
When the motor handpiece becomes dirty, wipe it off with a cotton cloth moistened with isopropyl alcohol.
CAUTION
• Do not lubricate the motor section of the motor handpiece.
• To clean the motor handpiece, do not use any solvent such as benzine and, thinner.
• Before mounting the lubricated contra angle head to the motor handpiece, wipe off extra oil. Stand on its end or
lean it in the proper position for gravity draining. Mount it after excess oil has been drained.
• Hold the contra angle head securely to prevent it from flying off by the pressure of the spray.
• Supply lubricant until it comes out of the handpiece head (for approx. 2 seconds).
• Do not use a spray can upside down. In such a case, only spray gas comes out and oil is not sprayed.

8. STERILIZATION

• Sterilize the contra angle head only.
• For the sterilization method, we recommend the autoclave sterilization method.
• Sterilization is required first time you use and after each patient as noted below.
Autoclaving
1) Brush the debris off the surface of the contra angle head, and wipe it off with a cotton cloth moistened with isopropyl
alcohol. Do not use metal brush.
2) Lubricate the head with the spray lubricant.
(Refer to the "P.14 7. (1) Lubricating Contra Angle Head.")
3) Insert the head into an autoclave pouch and seal it.
4) Autoclavable up to max. 135°C (275ºF).
ex.) Autoclave for 20 min. at 121°C (250ºF), or 15 min. at 132°C (270ºF).
5) Keep the contra angle head in the autoclave pouch to keep it clean until you use it.
* Sterilization at 121°C (250ºF) for more than 15 minutes is recommended by EN13060 or EN ISO17665-1.
Contra Angle Head
14
Spray Nozzle
Spray Lubricant
Fig. 12