Mounting Model F; Mounting Model G; Electrical Connections - E+E Elektronik EE23 Series Manual

Humidity/temperature transmitter
Hide thumbs Also See for EE23 Series:
Table of Contents

Advertisement

Available languages

Available languages

Electrical connections

V
mA
V
mA
24VDC/AC ±15%
terminal connection type EE23-FT
EE23 with connectors
description:
3
V+
2
4
GND
GND
1
5
RH
T
Euro-Norm
description:
NC
COM
NO
Euro-Norm
Mounting of Model F (wall mounting type with rear
cable outlet)
Transmitter EE23-xFx series are designed for wall mounting in clean room
applications. Connection cables do not disturb the cleaning process.
- working range sensor probe:
- working range electronics:
i
Mounting of Model G (wall mounting - for meteorology)
Transmitter EE23-xGx series are designed for outdoor applications.
- working range sensor probe:
- working range electronics:
For outdoor applications the transmitters must be equipped with a radiation
shield (refer to accessories). The shield causes forced ventilation which
prevents overheating of the sensor probe in direct sunlight. It also protects
the sensing probe against water of direct rainfall.
i
ON
terminal connection type EE23 with
alarm output module
T
5
Analog Ausgang /
RH
4
analogue output
GND
3
GND
2
Versorgung /
supply
V+
1
connection assignment:
connection assignment:
The transmitter must be mounted with the sensor probe pointed
downwards.
The sensor probe must point upwards and should be used in com-
bination with the radiation shield (refer to accessories).
NO
11
C
12
NC
13
V
mA
24VDC/AC ±15%
terminal connection type EE23-FPT
5
4
3
2
1
1
2
3
-40...60 degC (-40...140°F)
-40...60 degC (-40...140°F)
-40...60 degC (-40...140°F)
-40...60 degC (-40...140°F)
T
8
T
7
T-Sensor passiv /
T-Sensor passive
6
T
5
T
4
RH
Analog Ausgang /
analogue output
3
GND
2
GND
Versorgung /
supply
1
V+
15