Electrical Connection; Standard Connection; Connection Of The Plug Options - E+E Elektronik EE360 Operation Manual

Moisture in oil transmitter
Hide thumbs Also See for EE360:
Table of Contents

Advertisement

3.2

Electrical connection

The.electrical.installation.of.the.EE360.shall.be.performed.by.qualified.personnel.only..Observe.all.
applicable national and international requirements for the installation of electrical devices as well as for
power supply according to EN 61140, class III (EU) and class 2 supply (N. America).
3.2.1

Standard connection

8 - 35 V DC
12 - 30 V AC
V
mA
V
mA
Fig. 7
Electrical connection
Analogue outputs
Both.analogue.outputs.shall.be.configured.to.either.voltage.or.current..Measurands,.analogue.range.
and scaling are freely selectable. All settings can be performed via display and push buttons or using
the.EE-PCS.Product.Configuration.Software.
Error indication
The analogue outputs feature an error indication function according to NAMUR NE43. In the case of an
error the output signal will freeze at 21 mA or 11 V respectively.
note
By default the error indication is disabled.
When changing one of the analogue outputs from current to voltage and vice versa, the second output
will change automatically as well. The output scaling changes automatically if it is out of physical range
(i.e. 20 mA will be changed to 10 V instead of 20 V).
Check output scale after changing between voltage and current output.
3.2.2

Connection of the plug options

Option E4
power supply +
analogue output
Option E5
Operating instructions for EE360 Moisture in Oil
M16x1.5
V+
1
Supply
GND
2
GND
3
Analogue outputs
OUT 1
4
OUT 2
5
GND RS485
6
Modbus RTU
RS485 A (=D+)
7
(galvanically isolated)
RS485 B (=D-)
8
OUT 2
4
1
V+
n.c.
3
2
RS485 B
(=D-)
GND Output
GND
3
5
2
OUT 1
GND RS485
(shielded)
RS485 A
4
(=D+)
5
1
n.c.
9