Fill the KCl electrolyte reservoir
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Read the safety data sheet from the supplier
before bottles are filled or reagents are prepared. For laboratory use only. Make the hazard information
known in accordance with the local regulations of the user.
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
Note: To prepare 3M KCl electrolyte, refer to
Fill the KCl electrolyte reservoir with 3M KCl electrolyte as follows:
1. Put on the personal protective equipment identified in the safety data sheet (MSDS/SDS).
2. Turn the latch on the analytics panel to the unlocked position. Open the analytics panel.
3. Remove the lid from the KCl electrolyte reservoir. Refer to
4. Fill the reservoir (approximately 200 mL).
5. Install the lid.
6. From the front of the analytics panel, squeeze the KCl electrolyte tube with thumb and finger to
push the air bubbles up the tube to the reservoir. Refer to
When an air bubble is near the reservoir, use two hands to squeeze the tube on both sides of the
analytics panel to push the air bubble up.
7. Continue to squeeze the tube until the KCl electrolyte in the reference electrode is at the top of
the glass junction where the KCl electrolyte enters the electrode. Refer to
8. Close the analytics panel. Turn the latch on the analytics panel to the locked position.
Figure 5 Fill the KCl electrolyte reservoir
14 English