General Information - Hach NA5600 sc Na+ Installation Manual

Hide thumbs Also See for NA5600 sc Na+:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Specification
Interference phosphate 10 ppm Measurement interference is less than 0.1 ppb
Response time
Stabilization time
Calibration time
Calibration
Minimum detection limit
Auto calibration solution
Reactivation solution
3M KCl electrolyte
Conditioning solution
Concentration change from one channel
to another
0.1 ↔ 5 ppb
0.1 ↔ 50 ppb
0.1 ↔ 200 ppb
< 0.1 ↔ 1 ppb
0.1 ↔ 50 ppb

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer's website.
1
Experiment was done with ultra pure water (estimated at 50 ppt) and 1 ppb standard.
Table 3 Measurement specifications (continued)
Details
Refer to
Table
4.
Startup: 2 hours; Sample temperature variation: 10 minutes from 15 to 30 °C
(59 to 86 °F)
Use the optional heat exchanger when the temperature difference between
samples is more than 15 °C (27 °F).
50 minutes (typical)
Automatic calibration: known addition method; Manual calibration: 1 or 2 points
0.01 ppb
Approximately 500 mL of 10-ppm Sodium Chloride is used in 3 months with a 7-
day calibration interval.
Container: 0.5 L, HDPE with polypropylene caps
Approximately 500 mL of 0.5M Sodium Nitrate is used in 3 months with a 24-
hour reactivation interval.
Container: 0.5 L, HDPE with polypropylene caps
Approximately 200 mL of 3M KCl electrolyte is used in 3 months.
Container: 200 mL, polycarbon
Analyzers without cationic pump: Approximately 1 L of Diisopropylamine (DIPA)
is used in 2 months at 25 °C (77 °F) for a sample pH target of 11.2.
Approximately 1 L of DIPA is used in approximately 13 weeks at 25 °C (77 °F)
for a sample pH target of 10 to 10.5.
Analyzers with cationic pump: The usage rate of DIPA is dependent on the
selected Tgas/Twater ratio. With a ratio of 100% (i.e., the volume of sample is
equal to the volume of gas) the consumption of DIPA is approximately
90 mL/day.
Container: 1 L, glass with cap, 96 x 96.5 x 223.50 mm (3.78 x 3.80 x 8.80 in.)
Table 4 Average response times
T90% ≤ 10 minutes
Maximum temperature
1
difference (°C)
3
3
3
3
15
Time to accuracy 0.1 ppb or 5%
Up (minutes)
Down (minutes)
9
27
11
41
9
45
29
36
11
41
English 5