Annex; Technical Data; Analog Inputs - JUMO LOGOSCREEN 700 Brief Instructions

Highly-scalable paperless recorder
Hide thumbs Also See for LOGOSCREEN 700:
Table of Contents

Advertisement

6

Annex

6.1

Technical data

6.1.1

Analog inputs

General
Number
A/D converter
Sampling rate
Input filter
Galvanic isolation
Thermocouples
Designation
Type
Fe-CuNi
"L"
Fe-CuNi
"J"
Cu-CuNi
"U"
Cu-CuNi DIN
"T"
NiCr-Ni DIN
"K"
NiCr-CuNi
"E"
NiCrSi-NiSi
"N"
Pt10Rh-Pt
"S"
Pt13Rh-Pt
"R"
Pt30Rh-Pt6Rh
"B"
W5Re/W26Re
"C"
W3Re/W25Re
"D"
W5Re/W20Re
"A1"
Chromel-Copel
"L"
Chromel-Alumel
PLII (Platinel II)
Ambient temperature influ-
ence
Smallest measuring span
Measuring range start/end
Cold junction
Cold junction accuracy (in-
ternal)
Cold junction temperature
(external)
Base measuring range
a
Accuracy refers to the maximum measuring range. Small measuring spans lead to reduced linearization accuracy.
Max. 18 (see connection diagram)
24 bit delta-sigma
Up to 18 channels: 125 ms
Digital filter, 2nd order; filter constant can be set from 0 to 100.0 s
See "Galvanic isolation"
Standard
DIN 43710:1985-12
DIN EN 60584-1:2014
DIN 43710:1985-12
DIN EN 60584-1:2014
DIN EN 60584-1:2014
DIN EN 60584-1:2014
DIN EN 60584-1:2014
DIN EN 60584-1:2014
DIN EN 60584-1:2014
DIN EN 60584-1:2014
ASTM E230M-11
ASTM E1751M-09
GOST R 8.585-2001
GOST R 8.585-2001
GOST R 8.585-2001
≤ 100 ppm/K
Type L (Fe-CuNi), J, U, T, K, E, N, Chromel-Alumel, PLII: 100 K
Type S, R, B, C, D, A1, Chromel-Copel: 500 K
Freely programmable within the limits, in increments of 0.1 K
Internal (Pt100) or external (constant)
Option with 3 analog inputs (order codes 1 and 2): ± 1 K
Option with 6 analog inputs (order code 3): ± 2 K
-30 to +85 °C (adjustable)
20 to 70 mV
ITS
Measuring range
ITPS-68
-200 to +900 °C
ITS-90
-210 to +1200 °C
ITPS-68
-200 to +600 °C
ITS-90
-270 to +400 °C
ITS-90
-270 to +1372 °C
ITS-90
-270 to +1000 °C
ITS-90
-270 to +1300 °C
ITS-90
-50 to +1768 °C
ITS-90
-50 to +1768 °C
ITS-90
0 to 1820 °C
ITS-90
0 to 2315 °C
ITS-90
0 to 2315 °C
ITS-90
0 to 2500 °C
ITS-90
-200 to +800 °C
ITS-90
-270 to +1372 °C
0 to 1395 °C
6 Annex
a
Accuracy
≤ 0.1 %
≤ 0.1 % from -100 °C
≤ 0.1 % from -100 °C
≤ 0.1 % from -150 °C
≤ 0.1 % from -80 °C
≤ 0.1 % from -80 °C
≤ 0.1 % from -80 °C
≤ 0.1 % from 100 °C
≤ 0.1 % from 100 °C
≤ 0.1 % from 600 °C
≤ 0.1 % from 500 °C
≤ 0.1 % from 500 °C
≤ 0.1 % from 500 °C
≤ 0.1 % from -80°C
≤ 0.1 % from -80 °C
≤ 0.1 %
75

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents