temperature.
Measurements taken from the surrounding tissue of the ear canal
generate lower readings and may result in misdiagnosis of fever.
To avoid an inaccurate measurement:
1. Switch on the thermometer by pressing the ON/OFF button 4 .
2. After one beep is heard (and the temperature scale icon is
flashing), straighten the ear canal by gently pulling the middle of
the ear back and up.
3. Place the probe 1 firmly into the ear canal, press the START
button 3 and keep the probe in the ear until the thermometer
beeps to signal the completion of the measurement.

4. Control Displays and Symbols

• All segments displayed 7 : Press the ON/OFF button 4 to
turn on the unit; all segments will be shown for 1 second.
• Ready for measurement 9 : When the unit is ready for
measurement, the «° C » or «° F » icon will keep flashing.
IR1DQ1-1
Probe LED light is activated for 10 seconds and the unit auto-
off after 60 seconds. Press the START button to measure.
• Measurement complete AT :The reading will be shown on the
display 2 with the «° C » or «° F » icon; the unit is ready for the
next measurement, when the «° C »or«° F »icon is flashing again.
• Out-of-ear temperature indicator AK : A crossed-ear-icon will
appear on the display 2 if the reading falls outside the range
32.0 - 42.2 °C (89.6 - 108.0 °F).
• Low battery indicator AL : When the unit is turned on, the
triangle icon will keep flashing to remind the user to replace the
battery.

5. Set Date, Time and Beeper functions

• After the new batteries are fitted, the year number flashes in the
display BN. You can set the year by pressing the START button
3. To confirm and then set the month, press the M button 6 .
• Press the START button 3 to set the month. Press the M