Respironics BiPAP Plus User Manual page 44

Table of Contents

Advertisement

p
ressure
Pressure Increments:
Pressure Stability:
<10.0 cm H
O (±0.5 cm H
2
≥10.0 cm H
O (±1.0 cm H
2
Measured in accordance with EN ISO 17510-1 @ 1/3, 2/3, and Pmax with BPM set to 10, 15,
and 20 BPM @ 20°C (±5°C), 50% RH (±5%), and an atmospheric pressure of 101.54 kPascals.
Maximum Flow: 35 LPM
Measured in accordance with EN ISO 17510-1 @ 1/3, 2/3, and Pmax with BPM set to 10, 15,
and 20 BPM @ 23°C (±2°C), 50% RH (±5%), and an atmospheric pressure of 101.54 kPascals.
C
a
ontrol
CCuraCy
p
arameter
IPAP
EPAP
Ramp Duration
Rise Time
* Dynamic pressure accuracy is measured at the patient end of the circuit with a Whisper
Swivel II and varying flow conditions.
** Rise time settings are adjustable by the patient and the provider from 0 (off) to 3. Rise time
varies from 150 msec to 400msec dependent on patient setting and the pressure differential
from IPAP to EPAP. The rise time settings of "off" and "1" are functional only when the
pressure change between IPAP and EPAP is ≤ 8 cm H
there is little difference between settings. The rise time setting should be adjusted according to
the patient comfort level.
*** Measured at the patient end of the circuit with a Whisper Swivel II exhalation device and no
patient flow.
**** M easured in accordance with EN ISO 17510-1 @ 1/3, 2/3, and Pmax with BPM set to
10, 15, and 20 BPM @ 20°C (±5°C), 50% RH (±5%), and an atmospheric pressure of
101.54 kPascals.
d
Isposal
Dispose of the device in accordance with local regulations.
8-2
O)
2
O)
2
r
ange
4 to 25 cm H
O
2
4 to 25 cm H
O
2
0 to 45 minutes
0 to 3**
4.0 to 25.0 cm H
O (in 0.5 cm H
2
a
CCuraCy
± 1.5 cm H
O from static setpoint*/****
2
± 1.5 cm H
O from static setpoint*/****
2
± 10% of setting
± 25%***
O. At pressure differences > 8 cm H
2
B
pap p
I
O increments)
2
O,
2
m s
u
m
lus
erIes
ser
anual