Respironics BiPAP Plus User Manual page 32

Table of Contents

Advertisement

a
s
lert
ummary
The following table summarizes the high and medium priority alerts.
a
a
lert
udIBle
I
ndICator
System Error
• •
• •
Mask Leak
Auto Off
None
Patient
None
Reminder
5-2
t
aBle
v
d
Isual
evICe
I
a
ndICator
CtIon
Backlights:
The device
◊◊ ◊◊ ◊◊
enters the
"Safe state"
Screen displays
in which the
the following:
device power
remains on,
Service
but the airflow
Required
is disabled.
The device
Backlights:
continues to
operate.
The words
Mask Alert
flash on the
Active Display
screen.
The words Auto
The airflow
Off flash on the
shuts off and
Standby screen,
the device
indicating
enters the
that a mask
Standby state
disconnect has
approximately
been detected.
45-60 seconds
after detection.
The device
Backlights:
continues to
operate.
Screen displays
a customized
reminder
message
whenever
the device
transitions from
the Active state
to the Standby
state.
p
ossIBle
C
ause
Device failure
Press any of the display screen buttons
to silence the alert. Remove the power
supply cord from the device to remove
power. Plug the cord back into the device's
power inlet to restore power. If the alert
continues to occur, contact your home
care provider.
The breathing
Check your breathing circuit connections
circuit is
and reconnect the tubing if it has come
disconnected
loose. Press any of the display buttons to
or there is a
clear the alarm. Make sure your mask is on
large air leak.
properly before you press the
to restart the airflow.
If the alert continues to occur, disable
the mask alert setting following the
instructions in Chapter 4, and continue
your therapy overnight. Contact your
home care provider the next morning to
have your mask checked. You may need a
mask refitting.
The mask has
Put your mask back on and press the
been removed.
button to turn the airflow on and resume
therapy.
Your home care
Press any of the display buttons to
provider may
acknowledge the message and clear it. If
set a patient
you do not acknowledge the reminder,
reminder
it will disappear after 6 minutes, and
scheduled to
then reappear for three consecutive days
pop up at a
whenever the device transitions from the
particular time
Active state to the Standby state. If you
to remind you
still do not acknowledge the message,
to replace your
the reminder period will reset and the
mask, change
message will not reappear until the next
your filters, etc.
time the Patient Reminder setting expires.
B
pap p
I
p
a
atIent
CtIon
button
m s
u
m
lus
erIes
ser
anual