Correct Delay Control; Store And Recall Function Test - HP 8112A Operating, Programming And Servicing Manual

50 mhz programmable pulse generator
Table of Contents

Advertisement

Store and Recall
Function Test
Characteristics
P � o
� �ure
,
6. Set up the HP 81 16A Pulse/Function Generator as follows:
Output
Frequency
Amplitude
Offset
7. Verify that the scope display indicates a settling time of less than
200 µs as shown below.
'
Figure 8-21 . Correct Delay Control
Nine programmable locations for user preferred mode and parameter
settings
One for the standard (switch on) instrument mode and parameter set
(RCL 0).
One non accessible location for currently active mode and parameter
settings.
Press (sTotR cL)
Use right hand VERNIER key to obtain display
Press centre VERNIER key to recall the standard parameters.
Alter one of the parameters.
Press
Use right hand VERNIER key to obtain display UX, where X is
any number from 1 to 9.
Press left hand VERNIER key to store the altered parameter.
Squarewave
1 kHz
2.0 V
0.00 V
S e t t l i n g t i m e
. �
'
Testing Performance 8-29
1 00%
9 5 %
no