HP 8112A Operating, Programming And Servicing Manual page 143

50 mhz programmable pulse generator
Table of Contents

Advertisement

3. Verify that the result is:
W2 , D 1 , PER 1 . 0 MS , DTY 50 Y, , HIL 1 . 0 V , LOL 0 . 0 V , M 1 , CTO , T 1 , LO , CO , BUR 000 1 # ,
DEL 65 NS , WID 100 US , DBL 200 US , LEE 10 NS , TRE 10 NS
4. Use the following program statements to change some instrument
5. Verify that the settings are the same
6 . Using the oscilloscope connrm that th� HP 81 12A output has the
following form:
High-Level
Low-Level
settings and then re-read the current settings:
D IM B$ [ 1 6 1]
OUTPUT 712 ; 11PER 10 MS ,
DTY 10 Y, ,
W3 , HIL 1 . 5 V ,
DO"
OUTPUT 'f 12 ; " CST"
ENTER 712 ; B$
PRINT B$
following:
W3
DO
PER 10 . 0 MS
DTY 10 Y,
HIL 1 . 50 V
Pulses
Square
Period
10 ms
Duty Cycle
103
+ 1 .5 v
o.o v
Allocate controller memory to
receive second status string
Change settings
Request current settings from
HP 8112A
Read the HP 8112A settings
Display the HP 8112A settings
as
before, except for the
Testing Performance 8-3 1