Fast transitions

Verify that for the following HP 81 12A settings, the sampling scope
display indicates, the transition times are within the specifi e d
limits.
Record your results on a copy of the Test Record, specified limits
are given here and on the Test Record.
*
Under programming to 5.5 ns is permissible to meet this
Note
specification.
Slow transitions
Set up the HP 81 12A as follows:
Input Mode
Control Mode
PER
DEL
DTY
Transition
HIL
LOL
For the HP 81 12A to the settings in the table below, measure the
displayed leading edges
Verify that the transition times do not exceed the specifi e d limits.
Record your results on a copy of the Test Record, specified limits
are given here and on the Test Record.
The oscilloscope response is
Note
ms
the
provide faster response.
HP 8 1 12A
setting
Transition
PER
LEE
100
4.5
Fixed
ns
ns
*
100
6.5
Linear
ns
ns
NORM
Off
2 µs
65 ns
50%
Linear
+5.00 v
0.00 v
(
)
LEE
very
range. The HP 54503A or the HP 54100D oscilloscopes will
HP 81 12A
PER
LEE
TRE
2
500
500
µs
ns
500
100
100
µs
µs
2
500
500
ms
µs
5
999
999
µs
ms
5
1
1
ms
ms
ms
50 ms 10
10 ms 9.5
ms
Oscilloscope
TRE
Low Limit
High Limit
4.5
5.0
ns -
ns
(typical)
6.5
8.825
ns -
ns
(
and trailing edges
TRE
slow for measurements with PER in
Oscilloscope
Low Limit High Limit
473
527
ns
ns
ns
95
105
µs
µs
µs
475
525
µs
µs
µs
949
1.049
µs
µs
ms
0.95
1 .05
ms
ms
10.5
ms
ms
Testing Performance 8-15
)
.