Double Pulse
Performance Test
Note
Specifications
Equipment
Test Setup 1
8·6 Testing Performance
The DBL measurement is described as the time interval between the
two leading edges of the double pulse, recorded at 50% amplitude.
Range
20 ns to 950 ms
Accuracy
±5% of programmed value ±2 ns
Repeatability
Factor 4 better than accuracy
Max Jitter
0.2% of programmed value ± lOOps
Digitizing Oscilloscope with Accessory (HP 54121 T)
Counter (HP 5335A)
n,
Cable, 50
BNC to BNC, coaxial, 2 each (HP 8120- 1839).
20 dB attenuators
son
feedthrough termination
( Required if counter input impedance f 50
HP 8112A
Figure 8-3. Double Pulse Performance Test
n)
Oscilloscope (HP 5412xTI
20 dB Attenuator
with BNC/SMA Ada ptor