Apex Digital DOMUS 3 Instruction Manual page 10

Pump & mattress

Advertisement

English
3.
Low Pressure Indicator
When the mattress is initializing, the Low Pressure LED will illuminate until the appropriate
pressure is reached (according to the Pressure Adjustment Knob); this is normal.
Otherwise, the Low Pressure LED is a warning, indicating that the pressure in mattress is
unusually low. Check to ensure all connections are secured and correctly installed as per
the installation instructions.
NOTE: If the pressure level is consistently low, check for any leakage (tubes or air
hoses). If necessary, replace any damaged tubes or hoses or contact your
local qualified dealer for repair.
4.
Alternate / Static Switch
The ALTERNATE/STATIC switch selects between Alternate Pressure mode and Static
Pressure mode. With Alternate Pressure mode, alternating air cells are partially deflated
and inflated, avoiding prolonged pressure on any single point beneath the patient; this is to
prevent pressure ulcers. With Static Pressure mode, all of the air cells are equally inflated.
4.2 Emergency CPR Operations
CPR must be performed on a firm surface. Therefore, if
an emergency CPR situation occurs with the patient on
the mattress, the mattress must be quickly deflated.
This is done by quickly pulling on the CPR tag located at
the head of the mattress on the patient's right-side. The
quick connector found from the pump unit can also be
disconnected to hasten deflation.
5. CLEANING
It is important to follow the cleaning procedures before first-time use with any patient; this is to avoid
the possibility of transmission of pathogens between patients and caregivers.
Wipe down the pump unit with a damp cloth pre-soaked with a mild detergent. Avoid contact with
dust and proximity to dusty areas. Make sure that any cleaning agents you use will not harm or
corrode the plastic casing on the pump unit.
CAUTION: Do not immerse or soak the pump unit in liquids.
Wipe-down the mattress unit with a damp cloth pre-soaked with warm water containing a mild
detergent. Avoid dust and proximity to dusty areas. The cover may also be cleaned using sodium
hypochlorite diluted in water. All components should be air dried thoroughly before use.
8
Italiano
9. RISOLUZIONE DEI PROBLEMI
Problem
L'alimentazione non è
‧ Controllare se la spina è collegato alla rete.
acceso
‧ Controllare se l'alimentazione viene improvvisamente arrestato.
‧ Controllare se il CPR è sigillato.
Allarme è il
‧ Controllare se il collegamento tra il connettore del tubo aria pompa
(visivo)
sia collegato saldamente.
‧ Controllare se tutti i collegamenti dei tubi lungo materasso sono
garantiti..
‧ impostazione potrebbe essere insufficiente per il paziente, regolare
Il paziente sta finendo di
campo del comfort da 1 a 2 livelli più alti e attendere qualche minuto
pressione
per il miglior comfort.
‧ Controllare se tutti i bottoni a pressione o cinghie di materasso sono
Forma di materasso è
tutti saldamente fissati.
allentato
‧ Controllare se il materasso è fissato al telaio del letto con delle
cinghie
N aria prodotta da alcuni
‧ Questo è normale dal momento che è la modalità alternata. Prese
prese d'aria del tubo
d'aria, a turno, per produrre aria durante il loro tempo di ciclo
dell'aria connettore
Solution
101

Advertisement

loading