Removing/Replacing The Atf - CAIRE eQuinox PAOS 4000 Provider Technical Manual

Personal ambulatory oxygen system (paos)
Table of Contents

Advertisement

V. Removing/Replacing the ATF
1. To replace the ATF, Using a small utility knife, cut the two blue
tubing pieces connecting the ATF to the compressor as pictured
below in Figures 40 and 41 and discard.
2. Gently lift the ATF so as to allow access to the Drive Motor power
connector. Disconnect the Drive Motor power connector.
Personal Ambulatory Oxygen System Provider Technical Manual
Figure 41
Figure 42
Figure 43
3. Using a pair of small cutters, snip the end of the tie-wrap on the
soft tubing that connects the ATF to the Product Tank and discard
the tie-wrap.
4. Gently pull the soft tubing from the oxygen port on the ATF.
5. The ATF can now be removed from the unit by gently lifting
upward. To install a new ATF, follow steps 1-4 in reverse order.
Ensure to reinstall the tie-wrap on the soft tubing
connecting the ATF to the Product Tank. Also, prior
to attaching the blue tubing to connect to the
Note:
Compressor Assembly, use a cotton swab saturated
with Isopropyl Alcohol to swab the internal diameter
of the blue tubing. This will allow the blue rubber
tubing to slide over the aluminum tubing easier.
Figure 44
Figure 45
20771733 D 29