CAIRE Liberator 30 User Manual

CAIRE Liberator 30 User Manual

Oxygen reservoir
Hide thumbs Also See for Liberator 30:

Advertisement

Liberator
User Manual
(English)

Advertisement

Table of Contents
loading

Summary of Contents for CAIRE Liberator 30

  • Page 1 Liberator User Manual (English)
  • Page 2: Symbols Glossary

    Liberator Symbols Glossary This device complies with the re- ISO 7000; Graphical symbols for use on equip- quirements of Directive 2010/35/EU ment—Index and synopsis concerning medical devices. It bears Operating temperature limitation of the pi marking as shown. these units is 10°C to 40°C. Storage ADR: European Agreement concerning the Inter- temperature limitation range is -40°C national Carriage of Dangerous Goods by Road...
  • Page 3: Specifications

    12.5 mm in diameter and ingress of vertically dripping water. • Equipment not suitable for use in the presence of flammable mixtures Product Specifications Liberator 20 Liberator 30 Liberator 37 Liberator 45 Liberator 60 LOX Capacity 21.0 L 31.0 L...
  • Page 4: Warning Information

    Liberator Warning Information Smoking while wearing an oxygen can- nula can cause facial burns and possibly Important: Read this manual thoroughly before result in death. operating the Liberator. Removing the cannula and placing it RX Only. on clothing, bedding, sofas, or other cushion material will cause a flash fire WARNING: THIS DEVICE IS NOT IN- when exposed to a cigarette, heat source,...
  • Page 5 Liberator Introduction Controls 1. Gen 4 Meter Liquid Level Gauge The liquid oxygen system includes the Libera- tor and a portable unit, which provides you with 2. Flow Control Knob supplementary oxygen as prescribed by your physi- 3. DISS Connection cian.
  • Page 6: Basic Operations

    Liberator Caution: Should there be any liquid Operating Instructions leakage from the portable after separat- ing the units, set the portable aside, 1. To verify the level of liquid oxygen in the unit, ensuring it remains vertical, leave the see page 9. room, and call your health care provider immediately.
  • Page 7 Liberator 2. Use the following chart as a guideline to the recommended tubing length. FLOW MAXIMUM (RECOMMENDED) SETTING TUBING LENGTH* (LPM) 20-psig 50-psig 100 Ft. (30.5 m) 100 Ft. (30.5 m) 100 Ft. (30.5 m) 75 Ft. (22.9 m) 50 Ft. (15.2 m) 50 Ft.
  • Page 8: Battery Care And Maintenance

    Liberator Battery Care and Maintenance • Depress button to display level. If level is displayed and Low Battery Indicator is not il- luminated, battery level is acceptable. Humidifier Bottle and Cannula are not included 8. Adjust your breathing cannula to the proper posi- tion to breathe comfortably.
  • Page 9: Troubleshooting

    Liberator Troubleshooting Issue Solution Inadequate Flow • Verify flow control knob is on correct flow rate setting • Verify flow control knob is not set in between flow rates. • Verify liquid oxygen is in unit • Verify if cannula is kinked or pinched •...
  • Page 10: Transport And Storage

    Liberator Cleaning Standard Accessories WARNING: CLEAN ONLY AFTER THE UNIT IS EMPTY. • Clean using a solution of mild dish washing detergent and water. • Apply cleaning solution directly to a lint-free cloth. Approved cleaners include HydroPure and HydroKlean. Do not spray cleaners Cannula directly on the Liberator.
  • Page 11 Liberator Safety Caution: Medical Electrical Equipment needs special precautions regarding Electromagnetic compatibility (EMC) and needs to be installed and put into service according to the EMC infor- mation provided in this manual. Caution: Portable and mobile radio frequency (RF) communications equipment can affect Medical Electrical Equipment.
  • Page 12 Liberator Table 3 Guidance and manufacturers declaration—electromagnetic immunity The Liberator is intended for use in the electromagnetic environment specified below. The customer or the user of the Liberator should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment—guidance Electrostatic...
  • Page 13 Liberator Table 4* Guidance and manufacturers declaration—electromagnetic immunity The Liberator is intended for use in the electromagnetic environment specified below. The customer or the user of the Liberator should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment—guidance Conducted RF...
  • Page 14 Liberator Notes 14 - ENG PN MN234 C | User Manual...
  • Page 15 Liberator Notes PN MN234 C | User Manual 15 - ENG...
  • Page 16 . c om CAIRE Inc. 2200 Airport Industrial Dr., Ste. 500 Ball Ground, GA 30107 U.S.A. Medical Product Service GmbH Borngasse 20 35619 Braunfels, Germany Medstar Importacaoe Exportacao Ltda...