Removing/Replacing The Product Tank - CAIRE eQuinox PAOS 4000 Provider Technical Manual

Personal ambulatory oxygen system (paos)
Table of Contents

Advertisement

VI. Removing/Replacing the Product
Tank
1. Using a pair of small cutters, snip the end of the tie-wrap on the
soft tubing that connects the ATF to the Product Tank and discard
the tie-wrap.
2. Gently pull the soft tubing from the oxygen port on the ATF.
Personal Ambulatory Oxygen System Provider Technical Manual
Figure 46
Figure 47
3. Using a pair of small cutters, snip the end of the tie-wrap on the
bottom of the clear rubber elbow that connects the Product Tank
to the aluminum tubing going to the flow tube on the Control
Board.
4. The Product Tank can now be removed by gently pulling upward.
To install a new Product Tank, repeat steps 1-3 in reverse.
Ensure to reinstall the tie-wrap on the soft tubing
connecting the ATF to the Product Tank as well as the
Note:
clear rubber elbow that connects to the aluminum
tubing going to the flow tube on the Control Board.
Figure 48