eQuinox Voice Message Alert Descriptions Table

Condition
Flow rate error > +/- 10%
30 minutes runtime remaining on battery
15 minutes runtime remaining on battery
Low battery
Change in continuous flow setting
Change in pulse flow setting
Patient's breath rate not detected
Loss of power to unit
Unit is powered on in Pulse Mode and is using
AC power supply.
Unit is powered on in Continuous Mode and
is using AC power supply.
Unit is powered on in Continuous Mode and
is using battery.
Unit is powered on in Pulse Mode and is using
battery.
While unit is running from battery, the AC
power supply is connected.
While unit is running from AC power with the
battery installed and the AC power supply is
removed.
When powering unit off.
Personal Ambulatory Oxygen System Provider Technical Manual

eQuinox Voice Message Alert Descriptions

Standard eQuinox Voice Message
"Low output flow. Check the cannula. "
"30 minutes remaining on battery. "
"15 minutes remaining on battery. "
"Low on battery. Please plug in external
power. "
"Continuous mode flow xx liter per minute. "
(xx = [0.5L, 1.0L, 1.5L, 2.0L, 2.5L, 3.0L])
"Pulse dose mode xx milliliter. "
(xx = [16mL, 32mL, 48mL, 64mL, 80mL, 96mL,
128mL, 160mL, 192mL])
"Cannot detect breathing. "
"Loss of power. Please plug in external power
or battery. "
"The device is running. External power
connected. Pulse dose mode xx milliliter. "
(xx = [16mL, 32mL, 48mL, 64mL, 80mL, 96mL,
128mL, 160mL, 192mL])
"The device is running. External power
connected. Continuous dose mode xx liter
per minute. "
(xx = [0.5L, 1.0L, 1.5L, 2.0L, 2.5L, 3.0L])
"The device is running. Using battery.
Continuous dose mode xx liter per minute. "
(xx = [0.5L, 1.0L, 1.5L, 2.0L, 2.5L, 3.0L])
"The device is running. Using battery. Pulse
dose mode xx milliliter. "
(xx = [16mL, 32mL, 48mL, 64mL, 80mL, 96mL,
128mL, 160mL, 192mL])
"External power connected. "
"Using battery. "
"The device is now turned off. "
Table 8
Message Repeat Interval
Continuous until condition is removed.
One instance
One instance
Message played every 29 seconds until loss of
power state.
One instance
One instance
One instance
Continuously until condition is removed.
After start-up beep, on instance.
After start-up beep, on instance.
After start-up beep, on instance.
After start-up beep, on instance.
One instance
One instance
One instance
20771733 D 20