4. Remove the aluminum intake tube from the black rubber elbow.
5. Using a small utility knife, cut the two blue tubing pieces
connected to the ATF pictured below in Figures 38 and 39 and
discard.
Ensure that after the tubing has been removed
Note:
that the intake and vacuum ports on the ATF are
immediately capped to prevent damaging the ATF.
Personal Ambulatory Oxygen System Provider Technical Manual
Figure 38
Figure 39
6. The Compressor Assembly can now be removed/replaced. To
replace the Compressor Assembly, repeat steps 1-5 in reverse
order.
When installing the new Compressor Assembly, prior
to attaching the blue tubing to connect to the ATF,
use a cotton swab saturated with Isopropyl Alcohol
Note:
to swab the internal diameter of the blue tubing. This
will allow the blue rubber tubing to slide over the
aluminum tubing easier.
Figure 40
20771733 D 28