Annual Maintenance Procedures; Remove And Replace 9-Volt Battery And Compressor Intake Filter - CAIRE eQuinox PAOS 4000 Provider Technical Manual

Personal ambulatory oxygen system (paos)
Table of Contents

Advertisement

Annual Maintenance
Procedures
The following section lists procedures that are necessary to maintain
the eQuinox. Service should only be performed by a qualified
technician. To perform periodic maintenance, the only tools that
should be necessary are:
• #1 Phillips Screwdriver
• 3/8" or 10mm Deep Well Socket and Ratchet
WARNING:
Disconnect all power supplies going to the unit prior
to performing the following steps
Remove and Replace 9-Voly Battery and
Compressor intake Filter
Replace the 9-volt battery when the unit beeps three times at the end
of power-on self-test, when voltage is less than 7.0 Volts, or during
annual PM. To replace the 9-volt battery, follow these steps:
1. Remove the rear cover by loosening the two Phillips-head screws.
NOTE: The two captive screws will not remove completely from
the rear panel.
Personal Ambulatory Oxygen System Provider Technical Manual
Figure 7
2. Place finger under recessed area and pull outward.
3. Once the cover is removed, replace the 9-Volt Battery by pushing
up slightly on the bottom of the battery and then pulling out.
Figure 8
Figure 9
Figure 10
20771733 D 14