Download Print this page
CAIRE AirSep VisionAire 5 Quick Reference Manual

CAIRE AirSep VisionAire 5 Quick Reference Manual

Compact stationary oxygen concentrators

Advertisement

Quick Links

Flow Rates
VisionAire 5:
1.0 LPM – 5.0 LPM
VisionAire 3:
0.5 LPM – 3.0 LPM
VisionAire 2:
1/8 LPM – 2.0 LPM or 0.10 LPM – 1.0 LPM
Weight
30 lb (13.6 kg)
Sound Level
45 dB(A)
Power Consumption
VisionAire 5:
290 Watts
VisionAire 3:
175 Watts
VisionAire 2:
200 Watts
ON/OFF
With the unit plugged in, press the power switch upward in the ON (I) position. The audible alarm will sound briefly and oxygen flow will
begin. If the unit has O2 Monitoring, the yellow light will remain on during warm-up for 4 minutes after power up. If the light remains on for
15 minutes, an intermittent alarm will sound.
To turn the unit off, press the power switch downward in the OFF (0) position. All lights will turn off and oxygen flow will cease.
Nasal Cannula and Tubing
Standard oxygen tubing and nasal cannulas can be used with the VisionAire. Maximum extension tubing length is 50 ft (15.2 m) in addition to
a 7 ft (2.1 m) nasal cannula.

Humidifier Bottle

A humidifier bottle may be used with the VisionAire at all flow settings. It sits on the stand on the upper back of the unit and is secured with
an elastic strap. The stand is also sized to accommodate larger pre-filled humidifier bottles.
Preventative Maintenance
CAIRE does not require preventative maintenance on the VisionAire. VisionAire models are maintenance-free due to their technologically
advanced design. You do not need to perform service on the unit as long as the concentration remains within specifications.
Air Flow and Placement
Ambient air is drawn into the VisionAire through the side handles. Exhaust air exits the VisionAire through the bottom of the concentrator.
All intake and exhaust locations should remain un-obstructed. Ensure the VisionAire is located at least 12 in (30.5 cm) away from all walls,
draperies, furniture, etc.
Product Filter
The product filter, which is located within the unit's product tank, is designed to last the life of the unit. There is no required or scheduled
replacement needed for this part.
AirSep
VisionAire
®
Compact Stationary Oxygen Concentrators
Q
R
uick
efeRence
G
uide

Advertisement

loading

Summary of Contents for CAIRE AirSep VisionAire 5

  • Page 1: Humidifier Bottle

    The stand is also sized to accommodate larger pre-filled humidifier bottles. Preventative Maintenance CAIRE does not require preventative maintenance on the VisionAire. VisionAire models are maintenance-free due to their technologically advanced design. You do not need to perform service on the unit as long as the concentration remains within specifications.
  • Page 2: Oxygen Concentration

    The VisionAire 2 will produce oxygen at 95% (+0.5/-3%) up to 1 LPM, and at 90% (+5.5/-3%) between 1 LPM and 2 LPM. CAIRE recommends testing the oxygen concentration upon delivery to the patient and at periodic intervals set by the equipment provider. The provider may check the concentration as frequently as required.