6.12.1 Baud Rate

6.12.2 Sample Period

6.12.3 Duplex Mode

6.12.4 Linefeed

info@Fluke-Direct.com
The baud rate is the first parameter in the menu. The baud rate setting determines the serial
communications transmission rate.
Flashes bAUd and then displays the setting
bAUd
Current baud rate
2400 b
The baud rate of the serial communications may be programmed to 300, 600, 1200, 2400, 4800,
or 9600 baud. Use "UP" or "DOWN" to change the baud rate value.
New baud rate
4800 b
Press "SET" to set the baud rate to the new value or "EXIT" to abort the operation and skip to
the next parameter in the menu.
The sample period is the next parameter in the serial interface parameter menu. The sample
period is the time period in seconds between temperature measurements transmitted from the
serial interface. If the sample rate is set to 5, the instrument transmits the current measurement
over the serial interface approximately every five seconds. The automatic sampling is disabled
with a sample period of 0.
Flashes SPEr and then displays the setting
SPEr
Current sample period (seconds)
SP= 1
Adjust the value with "UP" or "DOWN" and then use "SET" to set the sample rate to the dis-
played value. Press "EXIT" to continue without changes.
New sample period
SP= 60
The next parameter is the duplex mode. The duplex mode may be set to full duplex or half du-
plex. With full duplex any commands are executed but not echoed. The duplex mode parameter
is indicated by,
Flashes dUPL and then displays the setting
dUPL
Current duplex mode setting
d=FULL
Adjust the setting using "UP" or "DOWN" and pressing "SET".
New duplex mode setting
d=HALF
The final parameter in the serial interface menu is the linefeed mode. This parameter enables
(on) or disables (off) transmission of a linefeed character (LF, ASCII 10) after transmission of
any carriage-return. The linefeed parameter is indicated by,