9100S Dry-well Calibrator
User's Guide
7.11
7.11.1
26
HL
Flashes
Flashes the current value and then displays the value for
375
adjustment
Current High Limit setting
375
Press the u or d to adjust the setting.
New High Limit setting
300
To accept the new setting, press "SET". Press "EXIT" to continue without stor-
ing the new value.
S
Accept the new High Limit setting

Serial Interface Parameters

The serial RS-232 interface parameters menu is indicated by,
SErL
Serial RS-232 interface parameters menu
Press "SET" to enter the menu. The serial interface parameters menu contains
parameters which determine the operation of the serial interface. The parame-
ters in the menu are — baud rate, sample period, duplex mode, and linefeed.

Baud Rate

The baud rate is the first parameter in the menu. The baud rate setting deter-
mines the serial communications transmission rate.
The baud rate parameter is indicated by,
bAUd
Flashes
Current baud rate
2400b
The baud rate of the serial communications may be programmed to 300, 600,
1200, 2400 (default), 4800, or 9600 baud. Use the up or down arrows to change
the baud rate value.
New baud rate
4800 b
Press "SET" to accept the new setting or "EXIT" to abort the operation and
skip to the next parameter in the menu.
and then displays the setting
for one second and then displays the setting