Fluke 9009 User Manual
Hide thumbs Also See for 9009:

Advertisement

Quick Links

9009
Industrial Dual-Block Calibrator
User's Guide
February 2013 Rev. 1, 3/22
© 2013-2022 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.

Advertisement

Table of Contents
loading

Summary of Contents for Fluke 9009

  • Page 1 9009 Industrial Dual-Block Calibrator User’s Guide February 2013 Rev. 1, 3/22 © 2013-2022 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
  • Page 2 Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
  • Page 3: Table Of Contents

    Table of Contents Title Page Introduction....................Safety Information ..................Contact Fluke Calibration ................Service Information..................Specifications .................... Quick Start....................Unpack the Product ................... Set Up......................Parts and Controls..................Top Panel (Lid Open) ................Constant Temperature Block ..............Accessories ....................General Operation ..................
  • Page 4 9009 User’s Guide Calibration Parameters ................. 14 R0 ...................... 15 ALPHA ....................15 DELTA ....................15 BETA (Low Temp Only) ..............15 Digital Communication Interface ............... 15 Serial Communications ................15 Wiring ....................16 Serial Operation..................16 Interface Commands ................. 17 Calibrate a Single Probe ................
  • Page 5: Introduction

    Introduction The Fluke Calibration 9009 Industrial Dual-Block Calibrator (the Product, calibrator, or instrument) is a portable instrument/bench-top temperature calibrator. Use the Product to calibrate thermocouple and RTD temperature probes. The Product is small enough to use in the field, and accurate enough to use in the lab.
  • Page 6: Safety Information

    Product. Contact Fluke Calibration Fluke Corporation operates worldwide. For local contact information, go to our website: www.flukecal.com To register your product, view, print, or download the latest manual or manual supplement, go to our website.
  • Page 7: Service Information

    Service Information Service Information Contact an authorized Fluke Calibration Service Center if the Product needs calibration or repair during the warranty period. Please have Product information such as the purchase date and serial number ready when you schedule a repair.
  • Page 8: Parts And Controls

    9009 User’s Guide Parts and Controls This section describes the Product and its parts. Top Panel (Lid Open) See Table for the top panel parts. Table 1. Top Panel DUAL-WELL DRY-BLOCK CALIBRATOR –15°C to 350°C 9009 9009 BEFORE CLOSING LID: REMOVE ALL PROBES FROM TEMPERATURE WELLS.
  • Page 9: Constant Temperature Block

    Industrial Dual-Block Calibrator Parts and Controls Table 1. Top Panel (cont.) Number Description Number Description Hot Display. Both displays not only shows the set and actual temperatures but also shows various calibrator PEM. The power switch is located on functions, settings, and constants. The the PEM.
  • Page 10: Accessories

    9009 User’s Guide Accessories The inserts in this section are available for the Product. Inserts may be used with either the hot or cold temperature wells. Figure 1. Constant Temperature Block Calibration Wells Removable insert sleeves 12.7 mm (1/2 in) O.D. x 101.6 mm (4 in) deep...
  • Page 11: General Operation

    Industrial Dual-Block Calibrator General Operation General Operation Change Display Units The temperature units are factory defaulted to Celsius. To change to Fahrenheit or back to Celsius: 1. Push SET and UP simultaneously. The temperature display changes units. 1. Push SET three times to store the changes. Un= C 2.
  • Page 12: Temperature Set-Point

    9009 User’s Guide Temperature Set-point Set the temperature set-point to any value within the range and resolution as given in the specifications. Do not exceed the safe upper temperature limit of any device inserted into the well. To set the temperature, select the set-point memory and adjust the set-point value.
  • Page 13 Industrial Dual-Block Calibrator General Operation Figure 2. Controller Operation Flowchart...
  • Page 14: Temperature Scale Units

    9009 User’s Guide Temperature Scale Units Temperature Scale Units. Scan Set and enable the scan rate so that when the set-point is changed the Product heats or cools at a specified rate (degrees per minute) until it reaches the new set-point. With the scan disabled the Product heats or cools at the maximum possible rate.
  • Page 15: Heater Power

    Industrial Dual-Block Calibrator Secondary Menu Heater Power The temperature controller controls the temperature of the well by pulsing the heater on and off. The total power applied to the heater is determined by the duty cycle or the ratio of heater on time to the pulse cycle time.
  • Page 16: Controller Configuration

    9009 User’s Guide The proportional band width is set at the factory. The proportional band width is easily adjusted from the front panel. Set the width to discrete values in degrees C or F, which depends on the selected units.
  • Page 17: Serial Interface Parameters

    Industrial Dual-Block Calibrator Controller Configuration Serial Interface Parameters The serial RS-232 interface parameters menu is indicated by, Serial RS-232 interface parameters menu SErIAL The serial interface parameters menu contains parameters which determine the operation of the serial interface. These controls only apply to instruments fitted with the serial interface. The parameters in the menu are: BAUD rate, sample period, duplex mode, and linefeed.
  • Page 18: Duplex Mode

    9009 User’s Guide Duplex Mode The duplex mode can be set to full duplex or half duplex. With full duplex, any commands received by the calibrator with the serial interface will be immediately echoed or transmitted back to the device of origin.
  • Page 19: Alpha

    Industrial Dual-Block Calibrator Digital Communication Interface The calibration parameters R0, ALPHA, DELTA, and BETA (cold side only) characterize the resistance-temperature relationship of the platinum control sensor. These parameters may be adjusted by an experienced user to improve the accuracy of the calibrator. The calibration parameter name flashes on the display and the Product shows the current value.
  • Page 20: Wiring

    9009 User’s Guide Wiring The three-conductor jack for the serial port is located on the top of the Product. Figure shows the pin- out of this connector and suggested cable wiring. The TxD line on one side connects to the RxD line on the other and vice-versa.
  • Page 21: Interface Commands

    Industrial Dual-Block Calibrator Interface Commands Interface Commands Note When you send a command, preface the command with either a C: (cold side) or H: (hot side). If the C: or H: are left off, the returned value is for the hot side. The various commands to access calibrator functions through the digital interface are listed in this section (see Table 3).
  • Page 22 9009 User’s Guide Table 3. Controller Communications Commands Command Command Command Returned Acceptable Returned Description Example Values Format Example Display Temperature set: Read current set-point s[etpoint] set: 999.99 {C or F} 150.00 C Set current set-point s[etpoint]=n s=350 Instrument Range...
  • Page 23: Calibrate A Single Probe

    -25 to 25 parameter Set BETA calibration be[ta]=n be=0.342 parameter Miscellaneous Read firmware version *ver[sion] *ver ver.9999,9.99 ver.9009,1.21 number Read structure of all h[elp] list of commands commands Read ALL operating list of parameters parameters Legend: [] Optional Command data...
  • Page 24: Calibrator Characteristics

    9009 User’s Guide Calibrator Characteristics There is a temperature gradient vertically in the test well. The heater has been applied to the block in such a way as to compensate for nominal heat losses out of the top of the calibrator. However, actual heat losses will vary with design of the thermometer probes inserted into the calibrator.
  • Page 25: Calibration Procedure

    Industrial Dual-Block Calibrator Calibration Procedure Calibration Procedure Follow these steps for the calibration procedure: 1. Choose the set-points to use in the calibration of the R0, ALPHA, DELTA, and BETA parameters. These set-points are generally -15 °C, 0 °C, 60 °C, and 110 °C (5 °F, 32 °F, 140 °F, and 230 °F) but other set-points can be used.
  • Page 26: Compute R0 & Alpha

    9009 User’s Guide Compute R0 & ALPHA delta − delta − rzero − alpha − Where: delta is the new value of DELTA computed above. Compute BETA (Cold Side Only) − delta rzero beta − − )( ) )( )
  • Page 27: Accuracy And Repeatability

    Product. If there are any questions, see Service Information. • Before the use of any cleaning or decontamination method except those recommended by Fluke Calibration, check with an Authorized Service Center to be sure that the proposed method does not damage the Product.
  • Page 28 9009 User’s Guide Table 4. Troubleshooting Problem Possible Causes and Solutions Power the Product on and watch the display. If the first number shown is less than -0005-, the Incorrect temperature reading Product has been re-initialized. Do the Master Reset Sequence to initialize the system as described in the problem, Controller locks up.
  • Page 29 Industrial Dual-Block Calibrator Troubleshooting Problems, Possible Causes, and Solutions Table 4. Troubleshooting (cont.) Problem Possible Causes and Solutions Possible RF energy emission. With the Product stable, slowly rotate the Product. If no change occurs, the Product may need to be calibrated.
  • Page 30 9009 User’s Guide...

Table of Contents