Tests, Adjustments, and Troubleshooting
System Verification
Interpreting the Verification Results
Figure 2-11
Data selected in the Output area of the System Verification dialog box.
At the top of the printed output is the name of the device, the serial number of
the device, and the date tested.
Each S-parameter measurement result is printed with frequency tested, lower
and upper limit lines, the measured data, and the result of the test.
Figure 2-11 Example of Printed Tabular Verification Results
Figure 2-12
selected in the Output area of the System Verification dialog box. The printed
graphical results show the following:
- the name of the device measured
- the serial number of the device
- the parameters measured
2-28
shows an example of typical verification results with Print Tabular
shows an example of typical verification results with Print Graphs
Keysight M937xA PXIe Service Guide