Tests, Adjustments, and Troubleshooting
Before You Begin
Equipment
b
PNA
b
Compression test set
b
Dynamic accuracy test set
b
Frequency counter
b
Signal generator
USB Thermocouple Power
Sensor
ECal module, 3.5 mm
Optional: Printer, Mouse,
Keyboard
BNC cable
3.5 mm test cable
(-m- to -f-), (Qty 3)
3.5 mm test cable
(-m- to -m-)
GPIB controller cable
GPIB cable (Qty 2)
2-4
Critical Specifications
Test Instruments and Software
Published port match
specifications
None specified
None specified
Freq: 10 MHz to 20 GHz
Accuracy: ±0.5 ppm
CW Freq: 1.185 GHz
Freq: DC to 33 GHz
Range: –35 to +20 dBm
Freq: 300 kHz to 26.5 GHz
N/A
Cables
50 ohm, length ³60 cm
N/A
N/A
N/A
N/A
Recommended
Alternate Model
Model or Part
Number
Number
N5241A/42A, or
None
N5221A/22A, or
E8363C/64C
U3070AK01
None
U3020AD01
None
53151A
Any equivalent
Option 001
N5181A
E8257D
Option 520
U8485A
N8485A power sensor
Option 200
(for Adjustments, also
requires N8482A for
frequencies below 10
MHz) with
N1912A/13A/14A
power meter
N4691B
85052B
Any printer with Microsoft Windows 7 driver.
8120-1839
Any equivalent
8121-2111
Any equivalent
(60 cm), or
N4373-61604 (90
cm)
11500E
Any equivalent
Any equivalent
c
M9036-31301
10833A (1 m)
Any
Keysight M937xA PXIe Service Guide
a
Use
P
P
P
P, A
P
P, A
b
P
--
A
P
A
P, A
P