Keysight Technologies M937 A PXIe Series Service Manual page 45

Vector network analyzer modules, 300 khz to 4/6.5/9/14/20/26.5 ghz
Hide thumbs Also See for M937 A PXIe Series:
Table of Contents

Advertisement

Tests, Adjustments, and Troubleshooting
System Verification
Figure 2-9 System Verification Dialog Box
5. In the Select Calibration Kit box, click on the calibration kit or electronic
calibration module (ECal) being used. The corresponding verification kit to
use is selected for you and displayed in the Compatible Verification Kit(s)
box.
6. Under Outputs, click one of the following options. Refer to
-Print Tabular Data: Prints the verification data in tabular form which
-Print Graphs: Prints the verification data in graphical form. The graphical
-File Tabular Data: Writes the tabular data to a text file in the
- File Graphs: Saves a screen image in PNG format in the
For printed output, it is assumed that the printer has been tested and the
Windows driver is installed for the printer that is being used. The system
verification test prints to the printer that has been designated as the
default printer. (On the Windows Desktop display, click on Start > Devices
and Printers to verify the printer setup.
Keysight M937xA PXIe Service Guide
includes measured data and uncertainty limits. For an example, refer to
Figure 2-11 on page
2-28.
form includes measured data trace, factory supplied data trace, and
uncertainty limits. For an example, refer to
C:\Users\Public\Public Documents\Network Analyzer\SysVer\
directory.
C:\Users\Public\Public Documents\Network Analyzer\SysVer\
directory.
Figure
2-9.
Figure 2-12 on page
2-29.
2-25