Operation Manual EE23 Humidity/Temperature Sensor BA_EE23_B2_e // v1.0 // technical data are subject to change // 194768...
Page 2
E+E Elektronik Ges.m.b.H. doesn’t accept warranty and liability claims neither upon this publication nor in case of improper treatment of the described products. The document may contain technical inaccuracies and typographical errors. The content will be revised on a regular basis. These changes will be implemented in later versions. The described products can be improved and changed at any time without prior notice.
The operation manual may not be used for the purposes of competition without the written consent of E+E Elektronik® and may not be forwarded to third parties. Copies may be made for internal purposes. All information, technical data and diagrams included in these instructions are based on the information available at the time of writing.
Environmental Aspects Products from E+E Elektronik® are developed and manufactured observing of all relevant requirements with respect to environment protection. Please observe local regulations for the device disposal. For disposal, the individual components of the device must be separated according to local recycling regulations.
- flashes during normal operation - continuously on indicates sensor damage 4. Push-Buttons The S1 and S2 push buttons are used for the EE23 adjustment as well as for returning to the factory calibration (see Calibration / Adjustment). 5. Level / Hysteresis Potentiometers for setting the threshold / hysteresis (available with alarm module only)
(with alarm module only). visual feedback when depressing the S2 push button (Fig. 1) 3. SET Indication for EE23 with optional alarm output 4. Td / Tf Dew point temperature / frost point temperature 5.
--> Error category 1 = non-critical error, can be solved by the user flashing --> Error category 2 = critical error, return the device for service Blue LED constantly lit --> analogue output is set to voltage. Orange LED constantly lit --> analogue output is set to current. 3. Interface: USB-Interface for service use Operating manual EE23 humidity / temperature sensor...
Return the faulty unit for service Temperature measurement failure ERROR: 07 ERROR: 09 Humidity measurement failure ERROR: 10 1) not available with 0 - 1 V output 2) see “Cleaning Instructions” at www.epluse.com/ee23 Tab. 2 Description of error messages on the display Operating manual EE23 humidity / temperature sensor...
5. Place the front cover onto the middle part and close tight the enclosure using the four screws included in the scope of supply. Drilling for round holes: Drilling for long holes: Fig. 5 Mounting of metal enclosure Fig. 6 Mounting of polycarbonat enclosure Operating manual EE23 humidity / temperature sensor...
T1 Wall mount T2 Duct mount Mounting onto DIN rails EE23 with polycarbonate enclosure can be mounted also on DIN rails with the bracket HA010203 (not included in the scope of supply, see “spare parts / accessories”). Mounting the Sensing Probe Whenever possible place the entire probe inside the space to be monitored.
1) Two M16x1.5 cable glands (one of them mounted onto the enclosure) 2) Power supply + analogue outputs: cable connector, 5-poles, straight M12 (Lumberg RKC 5/7) 3) Analogue output: cable connector, 5-poles, straight M12 (Lumberg RKC 5/7) 4) AC power supply: cable connector, 3-poles, straight (7/8ʺ-16UN) Operating manual EE23 humidity / temperature sensor...
Page 13
The protection of the supply cable against excess current and short-circuit must be designated to a wire cross section of 0.8mm² (AWG 18) (6A fuse). The back cover and the middle part of the metal enclosure must be grounded during operation! Operating manual EE23 humidity / temperature sensor...
1-point adjustment shall be performed at 50 % RH and 20 °C (68 °F). 1-point adjustment leads to very good accuracy within the range of main interest at the cost of the accuracy beyond this range. Adjustment of EE23 Model T1, T2, T4 and T6 via Push Buttons on the Electronics Board 5.2.1 2-Point RH Adjustment Procedure / T Adjustment Procedure 1.
Press BUTTON S2 for min. 3 seconds to exit adjustment procedure without storing the adjusted value. In both cases the LED „D1“ and the symbol „CAL“ on the LC display are deactivated. 5.2.3 Return to Factory Calibration To return to RH or T factory calibration set first the jumper to RH or to T respectively. During normal measuring mode (i.e. not during the adjustment procedure. LED D1 shall be off, the display shall not show “CAL”) pressing BUTTON S1 and S2 together for min 5 seconds. The return to factory calibration is confirmed by a short flash of the LED „D1“. Operating manual EE23 humidity / temperature sensor...
Adjustment of the EE23 Model T5 via the USB Service Interface 1. Download and install the EE-PCS Product Configuration Software from www.epluse.com/configurator 2. Connect the USB service interface of EE23-T5 to the PC. 3. Start the EE-PCS software. 4. Select the desired adjustment mode and follow the instructions of the EE-PCS.
1) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement). 2) only for models T1, T2, T4 and T6. Operating manual EE23 humidity / temperature sensor...
Operating Range Humidity Sensor Temperature [degC] The gray area shows the allowed measurement range for the humidity sensor. Operation beyond this range does not destroy the sensing element, but the specified measurement accuracy cannot be guaranteed. Operating manual EE23 humidity / temperature sensor...