Frequency Range Upgrade from RF or Microwave to Millimeter Wave Frequency
Install W56 and W57 if Option MPB and LNP are not present
66.Refer to
Figure
13. Locate W56, semi rigid cable,
between Low Band switch A11 J3, and YTF Input (left side SMA connector on YTF). Torque to
10 inch-pounds.
Figure 13
RF Section
67.Locate W57, semi rigid cable,
the Front End assembly to A12 Output (right side SMA connector on YTF) as shown in
13. Torque to 10 inch-pounds.
68.Locate W2
N9020-20173
Torque to 10 inch-lbs.
Complete the Hardware Installation
69.Replace the right side chassis, and front frame assembly. If Option EXM External Mixing is
present, reconnect the semi rigid cable behind the front panel EXT MIXER connector.
70.Replace the power cord and power on the instrument to assure the instrument boots up. Some
auto align failures may occur. The instrument should sweep full span (49.99 GHz if Option
550). Turn on the 50 MHz calibrator signal and tune the instrument to 50 MHz. Verify the 50
MHz signal is present.
71.Shut down the instrument. Once the instrument is completely shut down remove the power
cord.
72.Replace the top brace, instrument cover, and feet.
24

N9010-20006

in the kit and attach this cable between A13 J9 on
in the kit and install between Attenuator A and the input connector.
N9010-20005
in the kit and attach this cable
Installation Note N9030-90066