Frequency Range Upgrade from RF or Microwave to Millimeter Wave Frequency
Instrument Software Installation
Upgrade the software to the latest revision, or reinstall the current version (if available) when the
customer does not want to upgrade. Software installation assures that after licensing to the new
frequency range, any software related changes to the instrument files will be performed by the
software installation.
To verify that the current version, press System, Show, System and look for the "Instrument S/W
Revision"
The latest revision of the X-Series Signal Analyzers software may be downloaded from
http://www.keysight.com/find/xseries_software.
Contact the division for removal of old frequency range license
1. Connect the instrument to the Keysight LAN. The division will need to log into the instrument
and remove the License for the old frequency range.
2. E-mail the division at SUPPORT,CMS (K-Sonoma.exgen1). Message Subject: Frequency Range
Upgrade In this e-mail provide:
— Instrument model number and serial number
— The order number from the License Entitlement Certificate
— The LAN address of the instrument
3. The division will e-mail you when the license removal process is complete.