Download Print this page

Keysight PXA N9030A Installation Note page 21

Signal analyzer frequency range upgrade from rf or microwave to millimeter wave frequency
Hide thumbs Also See for PXA N9030A:

Advertisement

Frequency Range Upgrade from RF or Microwave to Millimeter Wave Frequency
57.Refer to
Figure
10. Locate W51
port 1 and the other end to Attenuator B output. Torque to 10 inch-lbs.
Figure 10
W51 Cable Installation
58.Locate W8
N9020-20131
the Front End assembly A13A1J2. Refer also to
59.Locate W20
N9020-20152
assembly A13A1J4. Torque to 10 inch-lbs.
60.Locate W3
N9020-20072
assembly J701. Torque to 10 inch-lbs.
61.Reattach W33 cable 8121-1940 from the A16 Reference assembly J705 to the Front End
assembly A13A1J6. Torque to 10 inch-lbs.
Installation Note N9030-90066
N9020-20171
(1) in the kit and install one end to top switch
(2) in the kit and install between the A11 Low Band Switch J2 and
(3) in the kit and install between the YTO and the Front End
(4) in the kit and install from Attenuator A Cal In and A16 Reference
Figure
11. Torque to 10 inch-lbs.
21

Advertisement