Download Print this page

Keysight PXA N9030A Installation Note page 10

Signal analyzer frequency range upgrade from rf or microwave to millimeter wave frequency
Hide thumbs Also See for PXA N9030A:

Advertisement

Frequency Range Upgrade from RF or Microwave to Millimeter Wave Frequency
7. The signal analyzer will automatically consume the License File. (This may take a few minutes)
When the License File is consumed the Keysight License Manager will display a "Successful
License Installation" message as shown in
Figure 2
Successful License Installation
Alternate Installation Procedure
The License File can be manually installed over USB or LAN by placing the license file in the
following folder on the signal analyzer
C:\Program Files\Agilent\licensing
Verify the License Installation
1. Cycle power on the signal analyzer and wait until the analyzer boots to the measurement
application screen. The instrument will fail RF Alignment and the noise floor will appear
bumpy.
2. Press System, Show, System to display a list of installed options.
3. Verify that the installed option list contains the newly installed N9030A-544 or N9030A-550.
10
Figure
2.
Installation Note N9030-90066

Advertisement

loading