Frequency Range Upgrade from RF or Microwave to Millimeter Wave Frequency
54.Refer to
Figure
7. Position the switch/bracket assembly into the frame. Route W7 ribbon cable
between the switches. Notice the position of the other ribbon cable W55 and wire harness
W12. Secure the switch assembly with two 0515-1035 screws (1). Torque to 9 inch-lbs. Insert
W7 ribbon cable into Low Band Switch header. Ensure the locking tabs are engaged.
Figure 7
Cable Routing