c. IMPORTANT: In the Installation Complete prompt, select "No, I will restart my computer later" and then click
the "Finish" button.

Step 4: Install the Module

PXI hardware does not support "hot-swap" (changing modules while power is applied to the chassis) cap-
abilities. Before installing or removing a module to/from the chassis, power off the chassis to prevent damage
to the module.
This module can be used in a chassis with PXIe or PXI-H chassis slots.
Recommended best practices to ensure proper and safe module operating
conditions
Ensure proper chassis air flow is maintained
l
Select a chassis that provides thermal protection if fans become inoperable or forced air cooling is
l
obstructed
Use slot blockers (Keysight model Y1212A, 5 per kit) and EMC filler panels in empty module slots to
l
ensure proper operating temperatures. Keysight chassis
optimize module temperature performance and reliability of test.
Set chassis fans to high or auto. Do not disable fans.