Polarizer Assembly Removal And Replacement; Q-Switch Replacement; Figure 2-15: Polarizer Being Removed - NORTHROP GRUMMAN Iklwa II TEM00 Service Manual

Green laser
Table of Contents

Advertisement

Polarizer Assembly Removal and Replacement

Q-Switch Replacement

© 2019 Cutting Edge Optronics
1. Locate and remove the two mounting screws securing the polarizer assembly
to the optical bench.

Figure 2-15: Polarizer Being Removed

2. Lift the polarizer assembly out of the laser.
3. Install new polarizer assembly with the mounting hardware. The mounting
hardware self-aligns the polarizer assembly. Torque the #8-32 fasteners to
25 in-lbs.
CAUTION
Do not run the chiller or laser with the coolant hoses and RF cable
disconnected from the Q-switch
WARNING:
Flexible tubing can be easily damaged. Use caution when handling.
1. Disconnect the coaxial BNC RF cable to the RF driver.
2. Carefully remove the hard plumbing using two (2) 9/16" (14mm) wrenches.
See Figure 2-16.
Chapter 3: Tuning and Testing Procedures
26
Patara II and Iklwa II Service Manual

Advertisement