Remove Q-Switch from Beam Path

© 2019 Cutting Edge Optronics
WARNING:
Flexible tubing can be easily damaged. Use caution when handling.
1.
See coolant line purge (page 14).
2. Cut the cable-tie and remove BNC connector.
3. Mark both SwageLok pieces with permanent marker as reference for
reinstallation. Carefully remove the hard plumbing using two (2) 9/16"
(14mm) wrenches. See Figure 2-20.

Figure 2-20: Removal of the Q-Switch Hard Plumbing

4. Remove the three (3) #8-32 x 1 ¼" screws holding the Q-switch assembly to
the laser bench (See Figure 2-21, below.)

Figure 2-21: Location of Q-Switch Attachment Hardware

5. Remove the Q-switch assembly from the housing.